Factory certificate –  for selected sensors

(test on the basis of IEC 60770)

  • Service free of charge from the ifm group of companies
  • Multi-point factory certificate available as PDF document, also after the unit has been ordered
  • Languages: in two languages English/German
  • For selected products:
    Available for the following product series:
    • LDLxxx, LW27xx
    • PI27xx, PI28xx (manufactured as of 01/2018)
    • PG27xx, PG28xx (manufactured as of 01/2019)
    • PM1xxx
    • TA20x, TA22xx, TA25xx, TA28xx, TD22xx, TD25xx, TD28xx, TD29xx (manufactured as of 01/2018)
    • TA11xx, TA13xx, TCCxxx
    • SM6x2x, SM7x2x, SM8x2x
    • SDx500, SDx600 (issue only possible for air characteristics)
  • Entering the serial number S / N is sufficient to download the certificate. See the picture below.

Manufacturer's certificate EN10204-2.1 (order code ZC0001)

(according to DIN50049-2.1)

  • Certifies that the listed articles / article groups conform to the indications in the corresponding data sheet and have been subjected to a final test.
  • The document is bilingual (German / English).

Manufacturer's certificate EN10204-2.2 (order code ZC0002)

(according to DIN 50049-2.2)

  • Certifies that the listed articles / article groups conform to the indications in the corresponding data sheet and have been subjected to a 100 % final test.
  • Annexes: description of the 100 % tests
  • The document is bilingual (German / English).

The following documents are available for selected fluid sensor products:

Inspection certificate material EN10204-3.1 (order code ZC0003)

(according to DIN 50049-3.1)

  • Results of batch-related tests to prove that the requirements regarding the material characteristics and quality are met.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Only issued for fluid sensor products. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration protocol pressure (order code ZC0004)

  • Calibration certificate
  • 6 calibration points in steps of 20 % of the measuring range. Other calibration points possible on request.
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for pressure sensors. Please refer to the accessories of the required sensor to find out if this is possible.

DAkkS calibration protocol pressure sensors (order code ZC0005)

  • DKD/DAkkS certificate (DAkkS = Deutsche Akkreditierungsstelle)
  • 11 calibration points. Further calibration points possible on request.
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for pressure sensors. Please refer to the accessories of the required sensor to find out if this is possible.

DAkkS calibration temperature sensors 3-point (order code ZC0013)

  • DKD/DAkkS certificate (DAkkS = Deutsche Akkreditierungsstelle)
  • Comparative calibration in a bath containing the continuously stirred liquid
  • 3 calibration points (65-85-123°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

DAkkS calibration temperature sensors 5-point (order code ZC0014)

  • DKD/DAkkS certificate (DAkkS = Deutsche Akkreditierungsstelle)
  • Comparative calibration in a bath containing the continuously stirred liquid
  • 5 calibration points (20-65-85-100-123°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

DAkkS calibration temperature sensors N-point (order code ZC0015)

  • DKD/DAkkS certificate (DAkkS = Deutsche Akkreditierungsstelle)
  • Comparative calibration in a bath containing the continuously stirred liquid
  • Number (max. 4) and position of the calibration points according to the customer request (temperature range: -20°C...150°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration temperature sensors 3-point (order code ZC0016)

  • Calibration certificate
  • Comparative calibration in a bath containing the continuously stirred liquid
  • 3 calibration points (65-85-123°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration temperature sensors 5-point (order code ZC0017)

  • Calibration certificate
  • Comparative calibration in a bath containing the continuously stirred liquid
  • 5 calibration points (20-65-85-100-123°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration temperature sensors N-point (order code ZC0018)

  • Calibration certificate
  • Comparative calibration in a bath containing the continuously stirred liquid
  • Number (max. 4) and position of the calibration points according to the customer request (temperature range: -20°C...150°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

Cleaning for oxygen applications​ (order code ZC0019)

  • Cleaning for oxygen applications (free of oil and grease).
  • The cleaned units are given a special marking and packaging.
  • Applies to the specified article(s) and must be requested when ordering the article.
  • For each unit ZC0019 has to be ordered separately.
  • Only issued for fluid sensor products. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration protocol SD (order code ZC0020)

  • Calibration certificate
  • Comparative calibration with standard reference
  • Calibration at 6 flow values (depending on the nominal diameter)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for flow sensors. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration SM (order code ZC0052)

  • Calibration certificate
  • Comparative calibration with standard reference
  • Calibration at 3/4 flow values (depending on the nominal diameter)
  • The calibration is carried out in factory setting.
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for flow sensors. Please refer to the accessories of the required sensor to find out if this is possible.

ISO calibration SU (order code ZC0053)

  • Calibration certificate
  • Comparative calibration with standard reference
  • Calibration at 4 flow values (depending on the nominal diameter)
  • The calibration is carried out in factory setting.
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for flow sensors. Please refer to the accessories of the required sensor to find out if this is possible.

Factory calibration for infra-red temperature sensors (order code ZC0061)

  • Calibration certificate
  • Created according to the requirements of DIN EN ISO 9001:2008
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible when giving the serial number. The certificate will then be dated with the day of calibration.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

Recalibration for infra-red temperature sensor TW20xx (order code ZC0062)

  • Calibration certificate
  • Created according to the requirements of DIN EN ISO 9001:2008
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s).
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

Recalibration for infra-red temperature sensor TW2002 (order code ZC0063)

  • Calibration certificate
  • Created according to the requirements of DIN EN ISO 9001:2008
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s).
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

A2LA calibration temperature sensors 3-point​ (order code ZC0064)

  • Calibration certificate
  • Created according to A2LA scope number: 3416 - Malvern, PA / ILAC MRA
  • Comparative calibration in a bath containing the continuously stirred liquid
  • 3 calibration points (65-85-123°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

A2LA calibration temperature sensors 5-point (order code ZC0065) 

  • Calibration certificate
  • Created according to A2LA scope number: 3416 - Malvern, PA / ILAC MRA
  • Comparative calibration in a bath containing the continuously stirred liquid
  • 5 calibration points (20-65-85-100-123°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

A2LA calibration temperature sensors N-point​ (order code ZC0066)

  • Calibration certificate
  • Created according to A2LA scope number: 3416 - Malvern, PA / ILAC MRA
  • Comparative calibration in a bath containing the continuously stirred liquid
  • Number (max. 4) and position of the calibration points according to the customer request (temperature range: -20°C...200°C)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for temperature sensors. Please refer to the accessories of the required sensor to find out if this is possible.

Recalibration for oil particle counter LDP (order code ZC0069)

  • Calibration certificate
  • Created according to the requirements of DIN EN ISO 9001:2008
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s).
  • Only issued for oil particle counter. Please refer to the accessories of the required sensor to find out if this is possible.

DAkkS calibration flow sensors 6-point (order code ZC0075)

  • DKD/DAkkS certificate (DAkkS = Deutsche Akkreditierungsstelle)
  • Comparative calibration with standard reference
  • Calibration at 6 flow values (depending on the nominal diameter)
  • The calibration is carried out in factory setting using the analogue output (current). Alternatively visually via display or analogue output (voltage).
  • Recalibrations are only possible if subjected unit is returned. The unit shall be in a clean state, without adaptations (condition as supplied to the customer). The declaration of decontamination has to be included.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.
  • Only issued for flow sensors. Please refer to the accessories of the required sensor to find out if this is possible.

Inspection certificate material EN10204-2.1/2.2/3.1 (order code ZC0076)

(according to ISO10474)

  • Certifies that the listed articles / article groups conform to the indications in the corresponding data sheet.
  • Results of batch-related tests to prove that the requirements regarding the material characteristics and quality are met.
  • The document is bilingual (German / English).
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Only issued for level sensors. Please refer to the accessories of the required sensor to find out if this is possible.

INMETRO calibration protocol temperature sensors (order code ZBR001)

  • Calibration certificate
  • Created according to INMETRO regulations
  • The document is available in Portuguese.  
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.

INMETRO calibration protocol flow sensors (order code ZBR002)

  • Calibration certificate
  • Created according to INMETRO regulations
  • The document is available in Portuguese. 
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.

INMETRO calibration protocol pressure sensors (order code ZBR003)

  • Calibration certificate
  • Created according to INMETRO regulations
  • The document is available in Portuguese. 
  • Applies to the specified article(s) and must be requested when ordering the article.
  • Subsequent orders are only possible if subjected unit is returned.