LDL sensors: Installation guidelines

For proper operation, some guidelines must be followed.

Orientation

Preferable mounting is just before or in a vertical pipe with flow going up to ensure a full pipe. When installed in a horizontal pipe, install the sensor at a maximum of 45° from horizontal to avoid the influence of air bubbles or deposits.

The measurement channel of the LDL200 must be installed parallel to the direction of flow. An alignment mark is laser etched on the body of the sensor indicating the correct orientation in the pipe.

Pipe size / recommended fittings

 

Pipe size

LDL100

LDL200
1"
E43310

weld fitting
Not recommended *
1-1/2" Not recommended *
2"
E30130

weld fitting

E33208
1.5" triclamp fitting

E33209

2" triclamp fitting
2-1/2"
3"
4"
> 4"

* Due to the length of the tip and location of the measuring channel, LDL200 cannot be used for pipe size under 2”