- SA flow sensor
- Select products
Select the best SA sensor for your application
| Sensor Family | Flow Rate | Line Size | Medium Temperature | Accuracy | Points to Consider |
|---|---|---|---|---|---|
SA |
0.1…6150 gpm 1...27000 scfm |
1/2"…16" | -4…212 °F -20…100 °C Higher upon request |
~ ±10% | • Pressure ≤ 1450 psi • Avoid rapid medium temperature changes |
| Part no. | Probe length (mm) |
Process connection | Output |
|---|---|---|---|
| SA6010 | 27 | 1/2" NPT |
|
| SA5010 | 45 | M18 x 1.5 (adapter required) |
|
| SA4110 | 100 | 8 mm Ø (compression fitting required) |
|
| SA4310 | 200 | 8 mm Ø (compression fitting required) |
|
| SA2000 | 19 | G 1/2 BSPP | |
| SA5000 | 45 | M18 x 1.5 (adapter required) |
|
| SA4100 | 100 | 8 mm Ø (compression fitting required) |
|
| SA4300 | 200 | 8 mm Ø (compression fitting required) |
|
| SA6014 | 27 | 1/2" NPT |
|
| SA5014 | 45 | M18 x 1.5 (adapter required) |
|
| SA4114 | 100 | 8 mm Ø (compression fitting required) |
|
| SA4314 | 200 | 8 mm Ø (compression fitting required) |
|
| SA2004 | 19 | G 1/2 BSPP | |
| SA5004 | 45 | M18 x 1.5 (adapter required) |
|
| SA4104 | 100 | 8 mm Ø (compression fitting required) |
|
| SA4304 | 200 | 8 mm Ø (compression fitting required) |
| Line size (in) |
Flow rate of liquid (gpm) |
Flow rate of gas (cfm) |
|---|---|---|
| 2 | 96.5 | 430 |
| 3 | 216 | 965 |
| 4 | 386 | 1720 |
| 6 | 865 | 3860 |
| 8 | 1540 | 6850 |
| 10 | 2400 | 10700 |
| 12 | 3460 | 15500 |
| 14 | 4720 | 21000 |
| 16 | 6150 | 27400 |
In relative mode (with a 0...100% signal and display), measurable flow rates are approximately doubled.
Other solutions for water-based media...
| Sensor Family | Flow Rate | Line Size | Medium Temperature | Accuracy | Points to Consider |
|---|---|---|---|---|---|
SM |
5 ml/min (0.08 gpm)…160 gpm |
1/4"…2" | 14…158 °F -4...194 °F (units with TFT display) |
~ ±1% | • Conductivity ≥ 20 µS/cm • Totalizer function • Pressure ≤ 232 psi • Versions with TFT display have selectable units of measure |
SB |
0.1…50 gpm | 3/4"…1-1/2" | 14…212 °F -10…100 °C |
~ ±5% | • No straight pipe runs required • Response time ≤ 10 msec • Pressure ≤ 580 psi |
SV |
0.26…26.4 gpm | 1/2"…3/4" | 14…194 °F -10…90 °C |
~ ±2% | • Display flow and temperature • Water content ≥ 95% • Pressure rating ≤ 174 psi • Long straight pipe required |
SI |
3…100 cm/s | 1/8"…8" | -13…176 °F -25…80 °C |
~ ±10% | • Pressure < 4350 psi • Sanitary versions available • Avoid rapid medium temperature changes |
Other solutions for oil-based media...
| Sensor Family | Flow Rate | Line Size | Medium Temperature | Accuracy | Points to Consider |
|---|---|---|---|---|---|
SB |
0.1…50 gpm | 3/4"…1-1/2" | 14…212 °F -10…100 °C |
~ ±5% | • No straight pipe runs required • Response time ≤ 10 msec • Pressure ≤ 580 psi • Oil viscosity ≤ 46 cSt |
SU |
0…52.8 gpm | 1/2"…1" | 14…176 °F -10…80 °C |
~ ±5% | • Totalizer function • Pressure ≤ 232 psi • Oil viscosity ≤ 68 cSt |
![]() SI |
3…100 cm/s | 1/8"…8" | -13…176 °F -25…80 °C |
~ ±10% | • Pressure < 4350 psi • Sanitary versions available • Avoid rapid medium temperature changes |
Other solutions for gases...
| Sensor Family | Flow Rate | Line Size | Medium Temperature | Accuracy | Points to Consider |
|---|---|---|---|---|---|
![]() SD |
0…412 scfm | 1/2"…2" | 32…140 °F 0…60 °C |
~ ±3% | • Totalizer function • Pressure ≤ 232 psi • Monitor air, CO2, N2 and Ar |
SI |
3…100 cm/s | 1/8"…8" | -13…176 °F -25…80 °C |
~ ±10% | • Pressure < 4350 psi • Sanitary versions available • Avoid rapid medium temperature changes |