Service Center 800-441-8246
  1. Products
  2. Sensors
  3. Vibration sensors
  4. Diagnostic Edge controllers
  5. VSE100

Diagnostic electronics for vibration sensors

VSE100

DIAGNOSTIC ELECTRONICS

Diagnostic electronics for vibration sensors

VSE100

DIAGNOSTIC ELECTRONICS

  • For the reliable vibration monitoring of machines and equipment
  • Condition-based maintenance thanks to monitoring of imbalance, rolling element bearing or gear state
  • Inputs for the simultaneous detection of up to four vibration sensors
  • Two switching outputs for pre-alarm and main alarm
  • analog inputs for monitoring further measurement variables
Learn more
  • ce
  • ukca
  • cul
Product characteristics
Frequency range [Hz] 0.1...12000
Communication interface Ethernet
Application
Design parameter setting via PC software VES004
Application continuous vibration monitoring
Electrical data
Operating voltage tolerance [%] 20
Operating voltage [V] 24 DC; (when using the IEPE input: 24 V + 20%; IEPE = Integrated Electronics Piezo Electric)
Current consumption [mA] 200; (24 V)
Protection class III
Inputs / outputs
Total number of inputs and outputs 16; (configurable)
Number of inputs and outputs Number of digital inputs: 8; Number of analog inputs: 2; Number of dynamic inputs: 4; Number of digital outputs: 10; Number of analog outputs: 1
Inputs
Total number of inputs 10
Number of digital inputs 8; (configurable)
Input circuit of digital inputs HTL
Number of analog inputs 2
Analog input (current) [mA] 4...20
Resolution of analog input 12
Number of dynamic inputs 4
Dynamic input - signal 0...10 mA / IEPE / 4...20 mA
Dynamic input - resolution [bit] 16
Dynamic input - frequency range [Hz] 0,1...12000
Dynamic input - sampling rate [kSamples] 100
Outputs
Total number of outputs 10
Output signal switching signal; analog signal
Electrical design PNP
Number of digital outputs 10; (configurable)
Output function normally open / closed; (configurable)
Max. voltage drop switching output DC [V] 2
Permanent current rating of switching output DC [mA] 100
Number of analog outputs 1; (configurable)
Analog current output [mA] 0...22; (0...10 V)
Max. load [Ω] 500
Short-circuit protection yes
Type of short-circuit protection yes (non-latching)
Overload protection yes
Measuring/setting range
Frequency range [Hz] 0.1...12000
Interfaces
Communication interface Ethernet
Protocol TCP/IP
Operating conditions
Ambient temperature [°C] 0...70
Storage temperature [°C] 0...70
Protection IP 20
Tests / approvals
EMC
EN IEC 61000-6-2
EN IEC 61000-6-4
MTTF [years] 89
UL approval
Ta 70 °C
Enclosure type Type 1
voltage supply Limited Voltage/Current
Mechanical data
Weight [g] 460.6
Housing Plastic housing
Type of mounting rail; (TH35/EN60715)
Dimensions [mm] 114 x 50 x 105
Material PA
Data memory
History memory yes
Data memory buffered yes
Data memory type ring memory; FIFO
Real-time clock yes
Memory location internal
Memory interval min. 1 min
Memory size 881664 data records
Accessories
Accessories (optional)
cross-over Ethernet patch cable for the direct connection to the PC
Remarks
Pack quantity 1 pcs.
Electrical connection
Connection Combicon: ; Maximum cable length: 250 m