Service Center 800-441-8246
  1. Products
  2. Sensors
  3. Flow sensors / flow meters
  4. SM magmeters for water based media
  5. SM6601

Magnetic-inductive flow meter

SM6601

SMN12GGXFRKG/US-100

Magnetic-inductive flow meter

SM6601

SMN12GGXFRKG/US-100

  • Precise measurement of flow, consumption and medium temperature
  • High accuracy, repeatability and measurement dynamics
  • With switching output, analog output and pulse output
  • Clearly visible numerical display
  • Convenient operation via buttons or parameter setting via IO-Link
Learn more
  • dnv_gl
  • ce
  • ukca
  • cul
  • iolink
  • ec19352004
  • crn
Product characteristics
Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Measuring range
1.5...396 gph 0.03...6.6 gpm
Process connection threaded connection 1/2" NPT DN15
Application
System gold-plated contacts
Application Totalizer function; for industrial applications
Media Conductive liquids; water; water-based media
Note on media
conductivity: ≥ 20 µS/cm
viscosity: < 70 mm²/s (40 °C)
Medium temperature [°F] 14...158
Pressure rating [bar] 16
Pressure rating [psi] 232
MAWP (for applications according to CRN) [bar] 17.7
Electrical data
Operating voltage [V] 18...30 DC; (to SELV/PELV)
Current consumption [mA] 95; (24 V)
Min. insulation resistance [MΩ] 100; (500 V DC)
Protection class III
Reverse polarity protection yes
Power-on delay time [s] 5
Inputs / outputs
Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Inputs
Inputs counter reset
Outputs
Total number of outputs 2
Output signal switching signal; analog signal; pulse signal; IO-Link; (configurable)
Electrical design PNP/NPN
Number of digital outputs 2
Output function normally open / closed; (configurable)
Max. voltage drop switching output DC [V] 2
Permanent current rating of switching output DC [mA] 200
Number of analog outputs 1
Analog current output [mA] 4...20; (scalable)
Max. load [Ω] 500
Analog voltage output [V] 0...10; (scalable)
Min. load resistance [Ω] 2000
Pulse output flow rate meter
Short-circuit protection yes
Type of short-circuit protection yes (non-latching)
Overload protection yes
Measuring/setting range
Measuring range
1.5...396 gph 0.03...6.6 gpm
Display range
-475.5...475.5 gph -7.925...7.925 gpm
Resolution
0.5 gph 0.01 gpm
Set point SP
3.5...396.5 gph 0.06...6.6 gpm
Reset point rP
1.5...394 gph 0.03...6.57 gpm
Analog start point ASP
0...318 gph 0...5.3 gpm
Analog end point AEP
78...396 gph 1.3...6.6 gpm
In steps of
0.5 gph 0.01 gpm
Volumetric flow quantity monitoring
Pulse value 0.01...30 000 000 gal
Pulse length [s] 0,01...2
Temperature monitoring
Measuring range [°F] -4...176
Resolution [°F] 0.1
Set point SP [°F] -2.5...176
Reset point rP [°F] -3.5...175
Analog start point [°F] -4...140.5
Analog end point [°F] 31.5...176
In steps of [°F] 0.5
Accuracy / deviations
Flow monitoring
Accuracy (in the measuring range) ± (0,8 % MW + 0,5 % MEW)
Repeatability ± 0,2% MEW
Temperature monitoring
Accuracy [K] ± 2,5 (Q > 0,26 gpm)
Reaction times
Flow monitoring
Response time [s] 0.15; (dAP = 0, T19)
Delay time programmable dS, dr [s] 0...50
Damping process value dAP [s] 0...5
Temperature monitoring
Dynamic response T05 / T09 [s] T09 = 20 (Q > 0,26 gpm)
Software / programming
Parameter setting options Flow monitoring; quantity meter; Preset counter; Temperature monitoring; hysteresis / window; normally open / closed; switching logic; current/voltage/pulse output; Start-up delay; display can be deactivated; Display unit
Interfaces
Communication interface IO-Link
Transmission type COM2 (38,4 kBaud)
IO-Link revision 1.1
SDCI standard IEC 61131-9
Profiles Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis
SIO mode yes
Required master port class A
Process data analog 3
Process data binary 2
Min. process cycle time [ms] 5
Supported DeviceIDs
Type of operation DeviceID
default 570
Operating conditions
Ambient temperature [°F] 14...140
Storage temperature [°F] -13...176
Protection IP 67
Tests / approvals
EMC
DIN EN 60947-5-9
Shock resistance
DIN EN 68000-2-27 20 g (11 ms)
Vibration resistance
DIN EN 60068-2-6 5 g (10...2000 Hz)
MTTF [years] 145
Pressure equipment directive sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Mechanical data
Weight [g] 593.5
Material stainless steel (1.4404 / 316L); PBT-GF20; PC; FKM; TPE
Materials (wetted parts) stainless steel (1.4404 / 316L); PEEK; FKM
Process connection threaded connection 1/2" NPT DN15
Displays / operating elements
Display
Display unit 6 x LED, green (gpm, gph, gal, °F, 10³, 1000 x 10³)
Switching status 2 x LED, yellow
Measured values alphanumeric display, 4-digit
Programming alphanumeric display, 4-digit
Remarks
Remarks
MW = Measured value
MEW = Final value of the measuring range
Pack quantity 1 pcs.
Electrical connection
Connection Connector: 1 x M12; coding: A; Contacts: gold-plated
Diagrams and graphs
Pressure loss
dP Pressure loss
Q volumetric flow quantity