1. Homepage
  2. Products
  3. Process sensors
  4. Flow sensors / flow meters
  5. Systems for the consumption measurement of compressed air & industrial gases
  6. SDG070

Compressed air meter

SDG070

SDG2 1/2"METRIS PB DN65 SS

Compressed air meter

SDG070

SDG2 1/2"METRIS PB DN65 SS

  • Precise measurement of flow and consumption
  • High accuracy, repeatability and measurement dynamics
  • For detecting compressed air in industrial use
  • With switching output, analog output and pulse output
  • Clearly visible 4-digit LED display
{{ifmVariantName}}



  • ce
Technical details
Product characteristics
Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Measuring range [m³/h] 6.7...2000
Process connection flange DN65 according to:DIN 2448
Application
Application Totalizer function
Media compressed air
Note on media
air quality ISO 8573-1
class 141
class 344
Medium temperature [°C] 0...60
Pressure rating [bar] 16
Pressure rating [Mpa] 1.6
Electrical data
Operating voltage [V] 19...30 DC
Current consumption [mA] < 100
Protection class III
Reverse polarity protection yes
Power-on delay time [s] 0.5
Inputs / outputs
Number of inputs and outputs Number of digital outputs: 2; Number of analog outputs: 1
Outputs
Total number of outputs 2
Output signal switching signal; analog signal; pulse signal; (configurable)
Electrical design PNP
Number of digital outputs 2
Output function normally open / closed; (configurable)
Max. voltage drop switching output DC [V] 2
Permanent current rating of switching output DC [mA] 250; (per output)
Number of analog outputs 1
Analog current output [mA] 4...20; (scalable)
Max. load [Ω] 500
Pulse output consumed quantity meter
Short-circuit protection yes
Type of short-circuit protection yes (non-latching)
Overload protection yes
Measuring/setting range
Measuring range [m³/h] 6.7...2000
Display range [m³/h] 0...2400
Volumetric flow quantity monitoring
Pulse value 1 Nm³
Pulse length [s] 0,1
Accuracy / deviations
Flow monitoring
Accuracy (in the measuring range) ± (3 % MW + 0,3 % MEW) / ± (6 % MW + 0,6 % MEW); (class 141 /; class 344)
Reaction times
Flow monitoring
Response time [s] 0.1; (dAP = 0)
Operating conditions
Ambient temperature [°C] 0...60
Storage temperature [°C] -20...85
Max. relative air humidity [%] 90
Protection IP 65
Tests / approvals
EMC
EN 61000-4-2 ESD 4 kV CD / 8 kV AD
EN 61000-4-3 HF radiated 10 V/m
EN 61000-4-4 Burst 2 kV
EN 61000-4-6 HF conducted 10 V
Vibration resistance
DIN IEC 68-2-6 5 g (55...2000 Hz)
MTTF [years] 214
Mechanical data
Weight [g] 9778
Material PBT-GF20; PC; PC; stainless steel (1.4301 / 304); FKM; Pipe section: steel galvanized
Materials (wetted parts) stainless steel (1.4401 / 316); stainless steel (1.4301 / 304); ceramics glass passivated; PEEK; polyester; FKM; aluminum anodized; Pipe section: steel galvanized
Process connection flange DN65 according to:DIN 2448
Remarks
Remarks
MW = Measured value
MEW = Final value of the measuring range
Measuring, display and setting ranges refer to standard volume flow according to DIN ISO 2533.
For information about installation and operation please see the operating instructions.
Pack quantity 1 pcs.
Electrical connection
Connection Connector: 1 x M12
ifm efector, inc. • 1100 Atwater Drive • Malvern • PA 19355 — We reserve the right to make technical alterations without prior notice. — EN-US — SDG070-00 — 21.05.2010 —