Product characteristics |
Number of inputs and outputs |
Number of digital outputs: 1; Number of analogue outputs: 1 |
Measuring range |
-25...160 °C |
-13...320 °F |
|
Process connection |
Aseptoflex |
Installation length EL [mm] |
40 |
Application |
Special feature |
Gold-plated contacts |
Measuring element |
1 x Pt 1000 + 1 x NTC; (thermically coupled, with backup function (temperature measuring even if one element fails)) |
Media |
liquids and gases |
Pressure rating [bar] |
50 |
Minimum installation depth [mm] |
15 |
No dead space |
yes |
Electrical data |
Operating voltage [V] |
20...32 DC; (according to EN 50178 SELV/PELV) |
Current consumption [mA] |
23; (24 V) |
Protection class |
III |
Reverse polarity protection |
yes |
Power-on delay time [s] |
4 |
Integrated watchdog |
yes |
Inputs / outputs |
Number of inputs and outputs |
Number of digital outputs: 1; Number of analogue outputs: 1 |
Outputs |
Total number of outputs |
2 |
Output signal |
switching signal; analogue signal; (configurable) |
Electrical design |
PNP/NPN |
Number of digital outputs |
1 |
Output function |
normally open / normally closed / diagnostic signal; (parameterisable) |
Max. voltage drop switching output DC [V] |
2 |
Permanent current rating of switching output DC [mA] |
250 |
Diagnostic output |
diagnostic output for drift and fault monitoring |
Number of analogue outputs |
1 |
Analogue current output [mA] |
4...20 |
Max. load [Ω] |
(Ub - 10 V) x 50 |
Short-circuit protection |
yes |
Type of short-circuit protection |
pulsed |
Overload protection |
yes |
Measuring/setting range |
Probe length L [mm] |
40 |
Measuring range |
-25...160 °C |
-13...320 °F |
|
Drift warning |
|
Drift alarm |
|
In steps of |
|
Resolution |
Resolution of analogue output [K] |
0.05 |
Accuracy / deviations |
Precision analogue output [K] |
± 0,2 (-10...100°C); ± 0,3 (-25...-10/100...150°C); (probe completely inserted into the measured medium up to the sealing chamfer) |
Drift monitoring [K] |
± 0,2 (-10...100°C); ± 0,3 (-25...-10/100...150°C) *); (probe completely inserted into the measured medium up to the sealing chamfer) |
Temperature coefficient [% of the span / 10 K] |
< ± 0,01; (in case of deviation from the reference condition 25 ± 5 °C) |
Response times |
Dynamic response T05 / T09 [s] |
6 / 13 |
Software / programming |
Parameter setting options |
drift warning/drift alarm threshold; Fail-Safe; Display unit; scaling of the analogue output; redundancy switching; performance diagnostic output; switching logic; normally open / normally closed; programmable via EPS interface |
Operating conditions |
Ambient temperature [°C] |
-25...70 |
Note on ambient temperature |
electronics: -25...70 °C |
Process connection: -32...170 °C |
|
Storage temperature [°C] |
-40...85 |
Protection |
IP 67; IP 69K |
Tests / approvals |
EMC |
DIN EN 61000-4-2 ESD |
4 kV CD / 8 kV AD |
EN 61000-4-3 HF radiated |
10 V/m |
DIN EN 61000-4-4 Burst |
2 kV |
EN 61000-4-6 HF conducted |
10 V |
|
Shock resistance |
DIN IEC 68-2-27 |
50 g (11 ms) |
|
Vibration resistance |
DIN EN 60068-2-6 |
20 g (10...2000 Hz) |
|
MTTF [years] |
220 |
Note on approval |
Including free 5-point calibration certificate. |
|
Mechanical data |
Weight [g] |
0.456 |
Materials |
stainless steel (1.4404 / 316L); PEI; FKM |
Materials (wetted parts) |
stainless steel (1.4404 / 316L) |
Process connection |
Aseptoflex |
Surface characteristics Ra/Rz of the wetted parts |
Ra: < 0.6 |
Installation length EL [mm] |
40 |
Remarks |
Remarks |
referring to UL: for use on a low voltage circuit with overcurrent protection |
in accordance with UL873 tab. 28.1 or Imax = 100/Ub (Ub = voltage of the circuit) |
The 3-A qualification is only valid if adapters with 3-A qualification are used for installation. |
|
Pack quantity |
1 pcs. |
Electrical connection |
Connection |
Connector: 1 x M12; Contacts: gold-plated |