Product characteristics |
Process connection |
G 2 DN50 external thread |
Application |
Special feature |
Gold-plated contacts |
Media |
ultra-pure water; water; hydrous media |
Note on media |
hydrous media: for media with >10 % additives, the repeatability is the only available value |
|
Medium temperature |
|
Min. bursting pressure |
|
Pressure rating |
|
Vacuum resistance [mbar] |
-1000 |
Electrical data |
Operating voltage [V] |
18...32 DC; (to SELV/PELV) |
Current consumption [mA] |
< 75 |
Protection class |
III |
Reverse polarity protection |
yes |
Power-on delay time [s] |
5 |
Measuring principle |
ultrasonic |
Inputs |
Inputs |
counter reset |
Outputs |
Output signal |
switching signal; pulse signal; analogue signal; IO-Link; frequency signal; diagnostic signal; totaliser switching signal |
Electrical design |
PNP/NPN |
Output function |
normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] |
2 |
Permanent current rating of switching output DC [mA] |
100 |
Switching frequency DC [Hz] |
0...10000 |
Analogue current output [mA] |
4...20 |
Max. load [Ω] |
500 |
Pulse output |
flow rate meter |
Short-circuit protection |
yes |
Type of short-circuit protection |
pulsed |
Overload protection |
yes |
Measuring/setting range |
Measuring range |
5...1000 l/min |
0.3...60 m³/h |
79...15850 gph |
1.32...264.18 gpm |
|
Display range |
-1200...1200 l/min |
-72...72 m³/h |
-19020...19020 gph |
-317...317 gpm |
|
Resolution |
0.1 l/min |
0.001 m³/h |
1 gph |
0.01 gpm |
|
Set point SP |
10.5...1000 l/min |
0.63...60 m³/h |
166...15850 gph |
2.77...264.17 gpm |
|
Reset point rP |
5.3...994.8 l/min |
0.318...59.688 m³/h |
84...15768 gph |
1.4...262.8 gpm |
|
Analogue start point ASP |
-1000...800 l/min |
-60...48000 m³/h |
-15850...12680 gph |
-264.17...211.34 gpm |
|
Analogue end point AEP |
-800...1000 l/min |
-48000...60 m³/h |
-12680...15850 gph |
-211.34...264.17 gpm |
|
Low flow cut-off LFC |
5...50 l/min |
0.3...3 m³/h |
79...793 gph |
1.32...13.21 gpm |
|
Frequency end point, FEP |
200.6...1000 l/min |
12.037...60 m³/h |
3180...15850 gph |
53...264.17 gpm |
|
Frequency at the end point FRP [Hz] |
1...10000 |
Volumetric flow quantity monitoring |
Pulse length [s] |
0.002...2 |
Pulse value |
0.1...99990000 I; 0.026...26414563.515 gal |
Temperature monitoring |
Measuring range |
|
Display range |
-44...124 °C |
-47.2...255.2 °F |
|
Resolution [°C] |
0.1 |
Set point SP |
-19.6...100 °C |
-3.2...212 °F |
|
Reset point rP |
-20...99.6 °C |
-4...211.2 °F |
|
Analogue start point |
-20...76 °C |
-4...168.8 °F |
|
Analogue end point |
|
Frequency start point, FSP |
|
Frequency end point, FEP |
|
Frequency at the end point FRP [Hz] |
1...10000 |
Accuracy / deviations |
Flow monitoring |
Accuracy (in the measuring range) |
± (1,0 % MW + 0,5 % MEW) |
Repeatability |
± 0,2 % MEW |
Temperature monitoring |
Accuracy [K] |
± 2,5 (Q > 5 % MEW) |
Temperature coefficient [% of the span / 10 K] |
0,2 |
Response times |
Flow monitoring |
Response time [s] |
< 0.25; (dAP = 0, T09) |
Damping process value dAP [s] |
0...5 |
Temperature monitoring |
Dynamic response T05 / T09 [s] |
5,7 / 86 |
Software / programming |
Diagnostic functions |
direction of flow detection; signal quality |
Interfaces |
Communication interface |
IO-Link |
Transmission type |
COM2 (38,4 kBaud) |
IO-Link revision |
1.1.3 |
SDCI standard |
IEC 61131-9: 2013-07 |
Profiles |
Identification and Diagnosis (0x4000) |
Required master port type |
A |
Process data analogue |
3 |
Process data binary |
2 |
Min. process cycle time [ms] |
9.6 |
IO-Link process data (cyclical) |
function |
bit length |
totaliser |
32 |
Flow monitoring |
32 |
Temperature monitoring |
32 |
status |
4 |
Output 1 |
1 |
Output 2 |
1 |
|
Supported DeviceIDs |
Type of operation |
DeviceID |
Default |
1461 |
|
Operating conditions |
Ambient temperature [°C] |
-20...60 |
Storage temperature [°C] |
-25...80 |
Protection |
IP 67 |
Tests / approvals |
EMC |
|
Shock resistance |
DIN IEC 68-2-27 |
20 g (11ms) |
|
Vibration resistance |
DIN IEC 68-2-6 |
5 g (10...2000Hz) |
|
Pressure Equipment Directive |
can be used for group 2 fluids; group 1 fluids on request |
Mechanical data |
Weight [g] |
1173 |
Type of mounting |
inlet pipe length 5xDN; outlet pipe length 1xDN |
Materials |
housing: stainless steel (1.4404 / 316L); Display: PFA; Sealing Display: FKM; connector: POKAN |
Materials (wetted parts) |
Pipe section: stainless steel (1.4404 / 316L); Process connection sealing: Centellen Flat seal |
Process connection |
G 2 DN50 external thread |
Surface characteristics Ra/Rz of the wetted parts |
1.25 |
Displays / operating elements |
Display |
|
colour display 1,44", 128 x 128 pixels |
Switching function |
2 x LED, yellow |
diagnosis |
1 x LED, three-colour |
|
Accessories |
Items supplied |
Flat seal 2, Centellen |
package insert |
|
Remarks |
Remarks |
MW = measured value |
MEW = Final value of the measuring range |
pulse and totaliser signal are only available for one of the two outputs |
the accuracy indications are adhered to over the entire application area |
|
Pack quantity |
1 pcs. |
Electrical connection |
Connection |
Connector: 1 x M12; coding: A; Contacts: gold-plated |
Diagrams and graphs |
Note on pressure loss |

|
derating ambient temperature |

|