Product characteristics |
Number of inputs and outputs |
Number of digital outputs: 2; Number of analogue outputs: 1 |
Process connection |
G 3/4 DN20 flat seal |
Application |
Special feature |
Gold-plated contacts |
Media |
conductive liquids; water; hydrous media |
Note on media |
conductivity: ≥ 20 µS/cm |
viscosity: < 70 mm²/s (40 °C) |
|
Medium temperature [°C] |
-20...90 |
Pressure rating [bar] |
16 |
Electrical data |
Operating voltage [V] |
18...30 DC; (to SELV/PELV) |
Current consumption [mA] |
< 80 |
Protection class |
III |
Reverse polarity protection |
yes |
Power-on delay time [s] |
5 |
Inputs / outputs |
Number of inputs and outputs |
Number of digital outputs: 2; Number of analogue outputs: 1 |
Inputs |
Inputs |
counter reset |
Outputs |
Total number of outputs |
2 |
Output signal |
switching signal; analogue signal; pulse signal; IO-Link; frequency signal; (configurable) |
Electrical design |
PNP/NPN |
Number of digital outputs |
2 |
Output function |
normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] |
2 |
Permanent current rating of switching output DC [mA] |
100 |
Number of analogue outputs |
1 |
Analogue current output [mA] |
4...20; (scalable) |
Max. load [Ω] |
500 |
Pulse output |
flow rate meter |
Short-circuit protection |
yes |
Type of short-circuit protection |
pulsed |
Overload protection |
yes |
Measuring/setting range |
Measuring range |
0.1...75 l/min |
0.006...4.5 m³/h |
1.2...1190 gph |
0.02...19.82 gpm |
|
Display range |
-90...90 l/min |
-5.4...5.4 m³/h |
-1426.8...1426.8 gph |
-23.78...23.78 gpm |
|
Resolution |
0.1 l/min |
0.006 m³/h |
0.6 gph |
0.01 gpm |
|
Set point SP |
0.5...75 l/min |
0.03...4.5 m³/h |
8.4...1189 gph |
0.14...19.81 gpm |
|
Reset point rP |
0.1...74.6 l/min |
0.006...4.48 m³/h |
1.2...1183 gph |
0.03...19.71 gpm |
|
Analogue start point ASP |
0...59.9 l/min |
0...3.6 m³/h |
0...950 gph |
0...15.82 gpm |
|
Analogue end point AEP |
15.1...75 l/min |
0.9...4.5 m³/h |
240...1189 gph |
3.99...19.81 gpm |
|
Low flow cut-off LFC |
0.1...3.8 l/min |
0.006...0.23 m³/h |
1.8...59.4 gph |
0.03...0.99 gpm |
|
Frequency end point, FEP |
15.1...75 l/min |
0.9...4.5 m³/h |
240...1189 gph |
3.99...19.81 gpm |
|
Frequency at the end point FRP [Hz] |
1...10000 |
Volumetric flow quantity monitoring |
Pulse length [s] |
0.003...2 |
Pulse value |
0.01...99990000 I |
Temperature monitoring |
Measuring range [°C] |
-20...90 |
Display range [°C] |
-42...112 |
Resolution [°C] |
0.1 |
Set point SP [°C] |
-19.6...90 |
Reset point rP [°C] |
-20...89.6 |
Analogue start point [°C] |
-20...68 |
Analogue end point [°C] |
2...90 |
In steps of [°C] |
0.1 |
Accuracy / deviations |
Flow monitoring |
Accuracy (in the measuring range) |
± (0,8 % MW + 0,2 % MEW) |
Repeatability |
± 0,2 % MEW |
Temperature monitoring |
Accuracy [K] |
± 2,5 (Q > 5 % MEW) |
Response times |
Flow monitoring |
Start-up delay [s] |
0...50 |
Response time [s] |
< 0.25; (dAP = 0, T09) |
Damping process value dAP [s] |
0...5 |
Temperature monitoring |
Response time [s] |
15; (Q > 10 % MEW, T09) |
Software / programming |
Parameter setting options |
hysteresis / window; normally open / normally closed; switching logic; frequency output; current/pulse output; start-up delay; display can be deactivated; Display unit |
Interfaces |
Communication interface |
IO-Link |
Transmission type |
COM2 (38,4 kBaud) |
IO-Link revision |
1.1 |
SDCI standard |
IEC 61131-9 |
Profiles |
Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis |
SIO mode |
yes |
Required master port type |
A |
Process data analogue |
3 |
Process data binary |
2 |
Min. process cycle time [ms] |
6 |
Supported DeviceIDs |
Type of operation |
DeviceID |
Default |
955 |
|
Operating conditions |
Ambient temperature [°C] |
-20...60 |
Storage temperature [°C] |
-25...80 |
Protection |
IP 65; IP 67 |
Tests / approvals |
EMC |
|
CPA approval |
model number |
006MI |
accuracy class |
- |
maximum allowable error |
± 1,0 % FS |
Q (min) |
0,006 m³/h |
Q (t) |
- |
Q (max) |
4,5 m³/h |
|
Shock resistance |
DIN IEC 68-2-27 |
20 g (11 ms) |
|
Vibration resistance |
DIN IEC 68-2-6: |
5 g (10...2000 Hz) |
|
MTTF [years] |
114 |
UL approval |
UL Approval no. |
I014 |
File number UL |
E174189 |
|
Pressure Equipment Directive |
Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request |
Mechanical data |
Weight [g] |
798 |
Materials |
stainless steel (1.4408/316); stainless steel (1.4404 / 316L); PC; PBT+PC-GF30 |
Materials (wetted parts) |
stainless steel (1.4404 / 316L); PEEK; carbon fibre PEEK; EPDM; Centellen |
Process connection |
G 3/4 DN20 flat seal |
Displays / operating elements |
Display |
|
colour display 1,44", 128 x 128 pixels |
|
2 x LED, yellow |
|
Remarks |
Remarks |
MW = measured value |
MEW = Final value of the measuring range |
|
Pack quantity |
1 pcs. |
Electrical connection |
Connection |
Connector: 1 x M12; coding: A; Contacts: gold-plated |
Diagrams and graphs |
|

|
Pressure loss / volumetric flow quantity |
|