1. Homepage
  2. Products
  3. Process sensors
  4. Pressure sensors
  5. Use in process technology at high temperatures
  6. PI2205

Flush pressure sensor with display

PI-004-REZ01-MFRKG/US/ /P

PI2205

{{'OPTIONAL_INFO' | productAvailability | error}}

Dear customer,
Unfortunately, this data sheet is not available in the requested language. If you have any further questions, you can contact us by e-mail.

Best regards
Your ifm team

  • For use in high temperatures in the food and beverage industry
  • Robust stainless steel housing with hygienic front-flush design
  • With two-wire and three-wire connection technology
  • Reduced wiring complexity when operated as two-wire current loop
  • Integrated clamp diaphragm seal with stainless steel membrane
  • fcm
  • ce
  • fda
  • ehedg
  • eac
  • cul
  • iolink
  • ec19352004
  • 3a
Download
Technische Details
Application
Application hygienic systems
Media viscous media and liquids with suspended particles; liquids and gases
Medium temperature [°C] -25...200; (-25...160 °C; Ambient temperature > 65 °C)
Pressure rating
30 bar 435 psi 3000 kPa
Min. bursting pressure
50 bar 725 psi 5000 kPa
Type of pressure relative pressure
Electrical data
Min. insulation resistance [MΩ] 100; (500 V DC)
Protection class III
Reverse polarity protection yes
Integrated watchdog yes
2-wire
Operating voltage [V] 20...32 DC
Current consumption [mA] 3.6...21
Power-on delay time [s] 1
3-wire
Operating voltage [V] 18...32 DC
Current consumption [mA] < 45
Power-on delay time [s] 0.5
Inputs / outputs
Number of inputs and outputs Number of digital outputs: 2; Number of analogue outputs: 1
Outputs
Total number of outputs 2
Output signal switching signal; analogue signal; IO-Link; (configurable)
Electrical design PNP/NPN
Number of digital outputs 2
Output function normally open / normally closed; (parameterisable)
Number of analogue outputs 1
Analogue current output [mA] 4...20, invertible; (scalable)
Short-circuit protection yes
Type of short-circuit protection pulsed
Overload protection yes
2-wire
Max. load [Ω] 300
3-wire
Max. voltage drop switching output DC [V] 2
Permanent current rating of switching output DC [mA] 250
Switching frequency DC [Hz] 125
Max. load [Ω] (Ub - 10 V) / 20 mA
Measuring/setting range
Measuring range
-1...4 bar -14.5...58 psi -100...400 kPa
Set point SP
-0.99...4 bar -14.35...58 psi -99...400 kPa
Reset point rP
-1...3.99 bar -14.5...57.85 psi -100...399 kPa
Analogue start point
-1...3 bar -14.5...43.5 psi -100...300 kPa
Analogue end point
0...4 bar 0...58 psi 0...400 kPa
In steps of
0.005 bar 0.05 psi 0.5 kPa
Factory setting
SP1 = 1.000 bar rP1 = 0.920 bar
SP2 = 3.000 bar rP2 = 2.920 bar
ASP = 0.000 bar AEP = 4.000 bar
Accuracy / deviations
Switch point accuracy [% of the span] < ± 0,2; (Turn down 1:1)
Repeatability [% of the span] < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1)
Characteristics deviation [% of the span] < ± 0,2; (Turn down 1:1 linearity, incl. hysteresis and repeatability limit value setting to DIN 16086)
Linearity deviation [% of the span] < ± 0,15; (Turn down 1:1)
Hysteresis deviation [% of the span] < ± 0,15; (Turn down 1:1)
Long-term stability [% of the span] < ± 0,1; (Turn down 1:1; per year)
Temperature coefficient zero point [% of the span / 10 K] < ± 0,2; (0...200 °C)
Temperature coefficient span [% of the span / 10 K] < ± 0,15; (0...200 °C)
Response times
Damping for the switching output dAP [s] 0...30
Damping for the analogue output dAA [s] 0.01...99.99
2-wire
Step response time analogue output [ms] 45
3-wire
Min. response time of switching output (dAP) [ms] 3
Step response time analogue output [ms] 7
Interfaces
Communication interface IO-Link
Transmission type COM2 (38,4 kBaud)
IO-Link revision 1.0
SDCI standard IO-Link Communication Specification, Version 1.0, January 2009, Ordner No: 10.002
IO-Link device ID 104 d / 00 00 68 h
Profiles no profile
SIO mode yes
Required master port type A
Process data analogue 1
Process data binary 2
Min. process cycle time [ms] 2.3
Operating conditions
Ambient temperature [°C] -25...80
Note on ambient temperature
-25...60 °C
Medium temperature: > 160 °C
Storage temperature [°C] -25...100
Protection IP 67; IP 68; IP 69K
Tests / approvals
EMC
DIN EN 61000-6-2
DIN EN 61000-6-3
Shock resistance
DIN EN 60068-2-27 50 g (11 ms)
Vibration resistance
DIN EN 60068-2-6 20 g (10...2000 Hz)
MTTF [years] 149
Note on approval
Including free 6-point calibration certificate.
Mechanical data
Weight [g] 917
Materials stainless steel (1.4404 / 316L); stainless steel (1.4435 / 316L); PBT; PEI; PFA
Materials (wetted parts) stainless steel (1.4435 / 316L)
Min. pressure cycles 10 million
Process connection Clamp DN 38 / 1,5"
Surface characteristics Ra/Rz of the wetted parts < 0.38 / -
Fill fluid for pressure transfer medical white oil (suited for use in food applications, FDA compliant)
Diaphragm diameter [mm] 34
Displays / operating elements
Display
Display unit LED, green
switching status LED, yellow
function display alphanumeric display, 4-digit
measured values alphanumeric display, 4-digit
Display unit bar; psi; kPa; % of the span
Remarks
Pack quantity 1 pcs.
Electrical connection
Connector: 1 x M12; Contacts: gold-plated
ifm electronic gmbh • Friedrichstraße 1 • 45128 Essen — We reserve the right to make technical alterations without prior notice. — EN-GB — PI2205-01 — 29.10.2015 —